DF55 5-diff Hematology Analyzer
Paramters:25 reportable parameters: WBC, Neu#, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCC, P-LCR6 research parameter
- Model:
Paramters:25 reportable parameters: WBC, Neu#, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCC, P-LCR6 research parameter
Paramters:
25 reportable parameters: WBC, Neu#, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCC, P-LCR
6 research parameters: ALY#, ALY%, LIC#, LIC%, NRBC#, NRBC%
2D scattergrams for WBC differention
Principles:
Flow Cytometry(FCM)+ Tri-angle laser scatter+chemical staining method for WBC differention
Impedance method for WBC, RBC and PLT test
Cyanide free colorimetry for HGB test
Repeatbaility
WBC:<=2.5%(4.00-15.00*109/L )
RBC:<=1.5%(3.50-6.0*1012/L )
HGB:<=1.5%(110-180g/L)
MCV:<=1%(70-120fL)
PLT:<=6.0%(100-149*109/L )
<=4.0%(150-500*109/L )
Sample volume: 20μL
Storage: Up to 100,000 records
Linearity range
WBC:0.00-300*109/L
RBC: 0.00-8.50*1012/L
HGB: 0-250g/L
PLT: 0-3000*109/L
HCT:0.00-67%
Net weight
28kg
Dimension
364mm(W)*498mm(H)*431mm(D)
Model Name:DF55
5-diff Hematology Analyzer
Up to 60 tests per hour
25 reportable parameters, 3 histograms
Whole blood, capilary whole blood and pre-diluted modes
10.4 inches TFT Touch Screen